The LI-COR Odyssey CLx Model 9140 is a dual-channel near-infrared fluorescence imaging system designed for quantitative Western blot analysis and other fluorescence-based laboratory applications.

This unit, S/N CLX-1715, was professionally evaluated and serviced by LabCycle Partners. During the initial assessment, the DDSP detector-acquisition board failed to enumerate on the instrument’s internal PCI bus. The board and associated connectors were inspected and securely reseated, restoring normal communication and acquisition operation without requiring replacement parts.

Following service, the instrument successfully completed a two-channel preview acquisition using LI-COR Image Studio software. The scan head homed and rastered, both the 700 nm and 800 nm channels were exercised, and the detector chain returned a complete image.

This testing verifies fundamental motion, laser excitation, detection, and imaging operation. Calibration and validation using a LI-COR Verification Plate were not performed.

FEATURES

  • Dual-channel near-infrared fluorescence imaging

  • 700 nm and 800 nm detection channels

  • Simultaneous two-color imaging

  • 25 cm × 25 cm flatbed scanning surface

  • Adjustable resolution from 21 µm to 337 µm

  • Motorized focus adjustment from 0 to 4 mm

  • AutoScan capability for extended dynamic range

  • Ethernet-based instrument communication

  • Compatible with LI-COR Image Studio and LI-COR Acquisition Software

  • Designed for membranes, gels, microplates, slides, and tissue sections

  • Suitable for quantitative Western blot and In-Cell Western applications

  • Professionally serviced and verified functional

SPECIFICATIONS

  • Manufacturer: LI-COR, Inc.

  • Product Line: Odyssey CLx

  • Model: 9140

  • Serial Number: CLX-1715

  • Manufacture Date: February 2017

  • Product Type: Dual-Channel Near-Infrared Imaging System

  • Excitation Wavelengths: 685 nm and 785 nm

  • Detection Channels: 700 nm and 800 nm

  • Detectors: Silicon avalanche photodiodes

  • Scanning Surface: 25 cm × 25 cm

  • Resolution Settings: 21, 42, 84, 169, and 337 µm

  • Scan Speed: 5–40 cm/second

  • Focus Offset Range: 0–4 mm

  • Supported Formats: Membranes, gels, microplates, slides, and tissue sections

  • Supported Microplates: 96-well and 384-well

  • Network Connection: Ethernet RJ-45

  • Input Voltage: 100–240 VAC

  • Input Current: 4 A

  • Frequency: 50/60 Hz

  • Approximate Dimensions: 21" W × 24.5" D × 15" H

  • Approximate Weight: 72 lb / 33 kg

  • Required Vertical Clearance with Lid Open: 29"

  • Laser Classification: Class 1 Laser Product

  • Country of Manufacture: United States

INCLUDED

  • LI-COR Odyssey CLx Model 9140 Infrared Imaging System, S/N CLX-1715

Includes all accessories shown in the photographs.

Computer, monitor, software media or license, verification plate, silicone mat, roller, cables, and additional accessories are not included.

CONDITION

Used – Professionally Serviced and Verified Functional.

The instrument was serviced after its DDSP detector-acquisition board failed to enumerate during startup. The board and its connections were inspected and reseated, resolving the fault without requiring replacement parts.

Following service, the instrument successfully completed a two-channel preview acquisition using LI-COR Image Studio software. The scan head homed and rastered, both channels were exercised, and the detector chain returned a complete image.

Calibration, quantitative performance validation, and testing with a LI-COR Verification Plate were not performed. Calibration status is unknown. Professional verification is recommended before use in calibrated, validated, or regulated workflows.

Please review all photographs carefully for the exact cosmetic condition and configuration.

SHIPPING

This item requires freight shipping due to its size, weight, and sensitive components.

Freight shipping is available throughout the United States. Shipping costs are quoted separately based on destination and delivery requirements.

Local pickup is available in San Diego, California.

Please complete our Freight Shipping / Local Pickup Form before purchasing so we can determine the appropriate shipping arrangements and provide an accurate freight quote.

The LI-COR Odyssey CLx Model 9140 is a dual-channel near-infrared fluorescence imaging system designed for quantitative Western blot analysis and other fluorescence-based laboratory applications.

This unit, S/N CLX-1715, was professionally evaluated and serviced by LabCycle Partners. During the initial assessment, the DDSP detector-acquisition board failed to enumerate on the instrument’s internal PCI bus. The board and associated connectors were inspected and securely reseated, restoring normal communication and acquisition operation without requiring replacement parts.

Following service, the instrument successfully completed a two-channel preview acquisition using LI-COR Image Studio software. The scan head homed and rastered, both the 700 nm and 800 nm channels were exercised, and the detector chain returned a complete image.

This testing verifies fundamental motion, laser excitation, detection, and imaging operation. Calibration and validation using a LI-COR Verification Plate were not performed.

FEATURES

  • Dual-channel near-infrared fluorescence imaging

  • 700 nm and 800 nm detection channels

  • Simultaneous two-color imaging

  • 25 cm × 25 cm flatbed scanning surface

  • Adjustable resolution from 21 µm to 337 µm

  • Motorized focus adjustment from 0 to 4 mm

  • AutoScan capability for extended dynamic range

  • Ethernet-based instrument communication

  • Compatible with LI-COR Image Studio and LI-COR Acquisition Software

  • Designed for membranes, gels, microplates, slides, and tissue sections

  • Suitable for quantitative Western blot and In-Cell Western applications

  • Professionally serviced and verified functional

SPECIFICATIONS

  • Manufacturer: LI-COR, Inc.

  • Product Line: Odyssey CLx

  • Model: 9140

  • Serial Number: CLX-1715

  • Manufacture Date: February 2017

  • Product Type: Dual-Channel Near-Infrared Imaging System

  • Excitation Wavelengths: 685 nm and 785 nm

  • Detection Channels: 700 nm and 800 nm

  • Detectors: Silicon avalanche photodiodes

  • Scanning Surface: 25 cm × 25 cm

  • Resolution Settings: 21, 42, 84, 169, and 337 µm

  • Scan Speed: 5–40 cm/second

  • Focus Offset Range: 0–4 mm

  • Supported Formats: Membranes, gels, microplates, slides, and tissue sections

  • Supported Microplates: 96-well and 384-well

  • Network Connection: Ethernet RJ-45

  • Input Voltage: 100–240 VAC

  • Input Current: 4 A

  • Frequency: 50/60 Hz

  • Approximate Dimensions: 21" W × 24.5" D × 15" H

  • Approximate Weight: 72 lb / 33 kg

  • Required Vertical Clearance with Lid Open: 29"

  • Laser Classification: Class 1 Laser Product

  • Country of Manufacture: United States

INCLUDED

  • LI-COR Odyssey CLx Model 9140 Infrared Imaging System, S/N CLX-1715

Includes all accessories shown in the photographs.

Computer, monitor, software media or license, verification plate, silicone mat, roller, cables, and additional accessories are not included.

CONDITION

Used – Professionally Serviced and Verified Functional.

The instrument was serviced after its DDSP detector-acquisition board failed to enumerate during startup. The board and its connections were inspected and reseated, resolving the fault without requiring replacement parts.

Following service, the instrument successfully completed a two-channel preview acquisition using LI-COR Image Studio software. The scan head homed and rastered, both channels were exercised, and the detector chain returned a complete image.

Calibration, quantitative performance validation, and testing with a LI-COR Verification Plate were not performed. Calibration status is unknown. Professional verification is recommended before use in calibrated, validated, or regulated workflows.

Please review all photographs carefully for the exact cosmetic condition and configuration.

SHIPPING

This item requires freight shipping due to its size, weight, and sensitive components.

Freight shipping is available throughout the United States. Shipping costs are quoted separately based on destination and delivery requirements.

Local pickup is available in San Diego, California.

Please complete our Freight Shipping / Local Pickup Form before purchasing so we can determine the appropriate shipping arrangements and provide an accurate freight quote.